SYNGO DYNAMICS (2021-04-30)

Starting date:
April 30, 2021
Posting date:
May 20, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75623

Last updated: 2021-05-20

Affected Products 

SYNGO DYNAMICS

Reason

Multiple Syngo Dynamics VA40D feature enhancements and the resolution to various software defects provided with this software are listed below:                                                                                                                                                                                                                   â€¢ Check conflicts does not work consistently across all the servers
• Data mining tool does not populate the calcs which have more than onecontextand present inmultiple departments
• CVIM # (stress testing) Reporting crashed when continuously slickingthe report toggle button inthumbnail mini toolbar
• RetrieveStudies.pl does not include database name in connectionstrings
• LV vol d, MOD A4Cdoes not have all the required disk measurements andother requiredmeasurements in the library.
• If a measurement namehas a different case with that of DB, SR importstops at thatmeasurement
• VA20C additional fetus isgetting created when SR objects are sentkeeping the worksheet andreport open.
• Studies are disappearing fromdatabase
• VA20F fixup tool still completesfixup and merges study even ifclicking on "X" in upper corneronthe possible merge dialog box.

Affected products

SYNGO DYNAMICS

Lot or serial number

VA40D

Model or catalog number

VA40

Companies
Manufacturer
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY