Health product recall

Synchromed II - Programmable Pump

Starting date:
August 22, 2017
Posting date:
September 11, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-64492

Reason

The investigation report (IR) for Synchromed II pump corrosion defines the problem statement as the "Degradation of Synchromed II motor components due to corrosion, wear, and a combination of corrosion and wear can result in temporary, intermittent, and/or permanent motor stall." The root cause of failure is identified as the depletion of lubricant at the shaft/jewel bearing interface which initiates mechanical wear on shaft ends, resulting in wear particles that can mix with the remaining lubricant creating residue. Shaft wear results in temporary or permanent motor stall.

Affected products

Synchromed II - Programmable Pump

Lot or serial number

Not Applicable

Model or catalog number

  • 8637-20
  • 8637-40