Symbia T series Gamma Camera/CT Scanners System(2019-09-24)
- Starting date:
- September 24, 2019
- Posting date:
- October 25, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71355
Last updated:
Affected Products
- Symbia T series Gamma Camera/CT Scanners
- Symbia Intevo
- Symbia Evo Excel
- Intevo Bold
Reason
In a limited number of systems, the SK2 connector has double wires crimped to one terminal which may result in wires intermittently opening and coming off the pin terminal. There are two cables per system. This may result in loss of voltage to high voltage board located on the detectors. The effect of losing power is that no image can be made.
Affected products
A. Symbia T series Gamma Camera/CT Scanners
Lot or serial number
1149
Model or catalog number
10275010
Companies
- Manufacturer
-
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES
B. Symbia Intevo
Lot or serial number
- 2138
- 2241
Model or catalog number
- 10764803
- 10764804
Companies
- Manufacturer
-
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES
C. Symbia Evo Excel
Lot or serial number
1163
Model or catalog number
10910501
Companies
- Manufacturer
-
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES
D. Intevo Bold
Lot or serial number
- 1148
- 2138
- 2241
Model or catalog number
11007962
Companies
- Manufacturer
-
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES