Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73921


Last updated:

Affected Products

A. Symbia T Systems

B. Symbia S System

Reason

Siemens has identified that certain Symbia S or Symbia T systems may have unintended slow motion of the detectors in specific situations which could pose potential patient risk. This issue may occur when a detector is moving downwards from approximately the zero-degree position radially towards the center of the camera bore, and an emergency stop press, touch pad engagement or power loss (i.e. system shut down or power failure) results.

Affected products

A. Symbia T Systems

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 8717733
  • 10275007
  • 10275009
Companies
Manufacturer

Siemens Medical Solutions USA, Inc. Molecular Imaging

2501 North Barrington Road

Hoffman Estates

60192

Illinois

UNITED STATES



B. Symbia S System

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

8717741

Companies
Manufacturer

Siemens Medical Solutions USA, Inc. Molecular Imaging

2501 North Barrington Road

Hoffman Estates

60192

Illinois

UNITED STATES