This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Symbia Intevo

Starting date:
March 16, 2017
Posting date:
March 31, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62862

Reason

Siemens Medical Solutions Usa, Inc. was made aware by the supplier that a potentially defective component was used in the line connection box (LCB) for a limited number of production systems. The LCB breaker may trip unnecessarily during normal operating conditions. This field action will replace the affected component and increase the stability of the breaker.

Affected products

Symbia Intevo

Lot or serial number

  • 2107
  • 2110
  • 2111

Model or catalog number

  • 10764804

Companies

Manufacturer

Siemens Medical Solutions USA, Inc. Molecular Imaging

2501 North Barrington Road

Hoffman Estates

60192

Illinois

UNITED STATES