Health product recall

SureSight Obturator - Introducer Localization System

Starting date:
July 29, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60062

Affected Products

SureSight Obturator - Introducer Localization System

Reason

THERE HAVE BEEN REPORTS THAT THE LOCALIZING OBTURATOR IS BREAKING BEFORE AND DURING PROCEDURES AT THE GLUE JOINT BETWEEN THE BODY AND THE TIP. THESE BREAKAGES IN THE FIELD MAY RESULT IN HARM TO THE PATIENT AND THE NEED FOR ADDITIONAL INTERVENTION TO REMOVE THE BROKEN PRODUCT FROM THE PATIENT.

Affected products

SureSight Obturator - Introducer Localization System

Lot or serial number

More than 10 numbers, contact manufacturer.

 

Model or catalog number

ILS-0914-12-OB

ILS-0914-20-OB

Companies
Manufacturer

Hologic, Inc.

250 Campus Drive

Marlborough

01752

Massachusetts

UNITED STATES