Stryker Precision Thin Blade Attachments
Last updated
Summary
Product
Stryker Precision Thin Blade Attachments
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Command2 Microelectric Blades | Not applicable. | 2296-003-108 |
| Precision Thin | 23070017 | 2296-003-155 |
Issue
During a review of the complaint handling and risk assessment data (October, 2024) it was identified that the precision thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.
Recall start date: January 8, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies
| Stryker Instruments, A Division Of Stryker Corporation |
| 1941 Stryker Way, Portage, Michigan, United States, 49002 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76837
Get notified