Stryker Instruments System 8 and 9 Saws
Last updated
Summary
Product
Stryker Instruments System 8 and 9 Saws
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| System 8 Reciprocating Saw | 2409624823 1802305513 | 8206-000-000 |
| System 9 Sternum Saw | Not applicable. | 9207-000-000 |
| System 9 Reciprocating Saw | More than 10 numbers, contact manufacturer. | 9206-000-000 |
Issue
There is a potential for the compression spring (0038-419-000) inside the saw's drivetrain assembly to disengage during blade loading and unloading. A disengaged spring disables the blade collar from springing back to its original home position, potentially leading to blade retention failure when loading a blade.
Recall Start Date: March 14, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Dental
Health products - Medical devices - Orthopaedics
Health products - Medical devices - General and plastic surgery
Companies
Stryker Instruments, A Division Of Stryker Corporation
1941 Stryker Way, Portage, Michigan, United States, 49002
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77182