Health product recall

Stryker Instruments System 8 and 9 Saws

Last updated

Summary

Product
Stryker Instruments System 8 and 9 Saws
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

System 8 Reciprocating Saw

2409624823
1802305513

8206-000-000

System 9 Sternum Saw

Not applicable.

9207-000-000

System 9 Reciprocating Saw

More than 10 numbers, contact manufacturer.

9206-000-000

Issue

There is a potential for the compression spring (0038-419-000) inside the saw's drivetrain assembly to disengage during blade loading and unloading. A disengaged spring disables the blade collar from springing back to its original home position, potentially leading to blade retention failure when loading a blade. 

Recall Start Date: March 14, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Dental
Health products - Medical devices - Orthopaedics
Health products - Medical devices - General and plastic surgery
Companies

Stryker Instruments, A Division Of Stryker Corporation

1941 Stryker Way, Portage, Michigan, United States, 49002

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77182

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe