Stryker Harmonic Focus Shears + Adaptive Tissue Technology (Reprocessed)
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Stryker Harmonic Focus Shears + Adaptive Tissue Technology (Reprocessed) |
More than 10 numbers, contact manufacturer. |
HAR9F |
Issue
Stryker has received an increase in complaints for HAR9F continuous activation due to the min button becoming stuck in the on position. Stryker has investigated and identified the manufacturing issue which contributed to the continuous activation noted with these devices. As of August 31, 2025, device failures related to this issue represent 0.16% of distributed devices; however, there have been no reported serious injuries.
Additional information
Details
Stryker Sustainability Solutions
1810 West Drake Dr., Tempe, Arizona, United States, 85283
Get notified
Receive emails about new and updated recall and safety alerts.