Health product recall

StrykeFlow Suction/Irrigator

Last updated

Summary

Product
StrykeFlow Suction/Irrigator
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

StrykeFlow Suction/Irrigator

More than 10 numbers, contact manufacturer.

0250070500

StrykeFlow Suction/Irrigator

More than 10 numbers, contact manufacturer.

0250070520

Issue

It has been identified that due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the StrykeFlow 2 Suction Irrigator.

Recall start date: July 2, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies

Stryker Endoscopy, A Div. Of Stryker Corp. Dba Stryker Endoscopy

5900 Optical Court, San Jose, California, United States, 95138

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77732

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe