Stratus CS 200 Acute Care Diagnostics System (2019-03-19)
- Starting date:
- March 19, 2019
- Posting date:
- April 29, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69780
Last updated: 2019-04-30
Affected Products
Stratus CS 200 Acute Care Diagnostics System
Reason
While using a POCT-1A communication protocol, the following error codes will not be sent from the Stratus CS 200 to the LIS.
- CH1H
- CH1L
- CH1M
- CH1N
- CH2H
- CH3H
- CS1K
- CS1V
The error messages will show on the instrument display and printout with the exception of CS1v. CS1V will only be displayed on the instrument as it indicates communication with the lis is not working. The other communication protocols are not affected by this issue.
Affected products
Stratus CS 200 Acute Care Diagnostics System
Lot or serial number
13050088
13050165
Model or catalog number
10714016
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, P.O. Box 6101
Newark
19714
Delaware
UNITED STATES