Health product recall

Stratus CS 200 Acute Care Diagnostics System (2019-03-19)

Starting date:
March 19, 2019
Posting date:
April 29, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69780

Last updated: 2019-04-30

Affected Products

Stratus CS 200 Acute Care Diagnostics System

Reason

While using a POCT-1A communication protocol, the following error codes will not be sent from the Stratus CS 200 to the LIS.
- CH1H
- CH1L
- CH1M
- CH1N
- CH2H
- CH3H
- CS1K
- CS1V
The error messages will show on the instrument display and printout with the exception of CS1v. CS1V will only be displayed on the instrument as it indicates communication with the lis is not working. The other communication protocols are not affected by this issue.

Affected products

Stratus CS 200 Acute Care Diagnostics System

Lot or serial number

13050088
13050165

Model or catalog number

10714016

Companies
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, P.O. Box 6101
Newark
19714
Delaware
UNITED STATES