Health product recall

StrataMR Adjustment Tools and StrataMR Guider Tool (2018-10-25)

Starting date:
October 25, 2018
Posting date:
November 28, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68394

Affected products

  1. StrataMR Adjustment Tools
  2. StrataMR Guider Tool

Reason

Medtronic is initiating a voluntary correction because there is a modification to the StrataMR Adjustment Tool Set which adds a fourth component, the Guider Tool, in order to improve users' ability to reliably adjust StrataMR valves and mitigate the potential of valve rotor foot being positioned on top of an MRI Resistance Wall, which has the potential of causing a higher than intended flow resistance. There is no change to the design of the existing three tools of the StrataMR adjustment tool sets. The StrataMR valves and shunts have been removed from the market as of February 2017; therefore this correction is intended for use with patients who currently have the StrataMR valves implanted.

Affected products

A. StrataMR Adjustment Tools

Lot or serial number

All lots.

Model or catalog number

45905

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES