Health product recall

Sterizone VP4 Sterilizer (2020-02-19)



Starting date:
February 19, 2020
Posting date:
March 13, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72567


Last updated:

Affected Products

Sterizone VP4 Sterilizer

Reason

The current vaporization block, a component inside the Sterizone VP4 Sterilizer, can become occluded over time when used with Sterizone 125-280 solution. In the event of an occluded vaporization block, an audible alarm indicates that an incomplete sterilization cycle has occurred

Affected products

Sterizone VP4 Sterilizer

Lot or serial number

All lots.

Model or catalog number

44000-R

Companies
Manufacturer

TSO3 INC.

2505 DALTON AVE.

SAINTE FOY

G1P 3S5

Quebec

CANADA