Steris AMSCO Eagle 3000 Prevacuum Sealer (2019-10-23)
- Starting date:
- October 23, 2019
- Posting date:
- November 8, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71557
Last updated: 2019-11-08
Affected Products
Steris AMSCO Eagle 3000 Prevacuum Sealer
Reason
Steris has identified that a small number of preventive maintenance kits for the Eagle 3000 small sterilizer were assembled with a 100 psig safety valve instead of a 40 psig valve, potentially resulting in the 100 psig valves being installed on sterilizers in the field. Use of the incorrect valve could result in the improbable scenario of water and steam leaking from the unit's fittings, or accumulation of steam in the chamber. Steris has not received any reports of adverse events related to the issue described above.
Affected products
Steris AMSCO Eagle 3000 Prevacuum Sealer
Lot or serial number
All lots.
Model or catalog number
- AX00-XXX
- AX01-XXX-XXX
- AX73-350
- AX73-358
- AX73-XXX-XXX-XXX
- FF33-XXX-XXX-XX
- FF43-XXX-XXX-XX
- FF53-XXX-XXX-XX
- FF63-XXX-XXX-XX
- FF73-XXX-XXX-XX