Health product recall

Steris AMSCO Eagle 3000 Prevacuum Sealer (2019-10-23)

Starting date:
October 23, 2019
Posting date:
November 8, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71557



Last updated: 2019-11-08

Affected Products

Steris AMSCO Eagle 3000 Prevacuum Sealer

Reason

Steris has identified that a small number of preventive maintenance kits for the Eagle 3000 small sterilizer were assembled with a 100 psig safety valve instead of a 40 psig valve, potentially resulting in the 100 psig valves being installed on sterilizers in the field. Use of the incorrect valve could result in the improbable scenario of water and steam leaking from the unit's fittings, or accumulation of steam in the chamber. Steris has not received any reports of adverse events related to the issue described above.

Affected products

Steris AMSCO Eagle 3000 Prevacuum Sealer

Lot or serial number

All lots.

Model or catalog number
  • AX00-XXX
  • AX01-XXX-XXX
  • AX73-350
  • AX73-358
  • AX73-XXX-XXX-XXX
  • FF33-XXX-XXX-XX
  • FF43-XXX-XXX-XX
  • FF53-XXX-XXX-XX
  • FF63-XXX-XXX-XX
  • FF73-XXX-XXX-XX
Companies
Manufacturer

Steris Mexico

790 Avante Ave., Industrial Park

Guadalupe, Nuevo Leon

67190

MEXICO