This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Steris 4085 General Surgical Table (2014-07-15)

Starting date:
July 15, 2014
Posting date:
August 1, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40817

Recalled Products.

Steris 4085 General Surgical Table

Reason

Steris engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.

Affected products

Steris 4085 General Surgical Table

Lot or serial number

0418009043 to 0417114082

Model or catalog number

ST01XXXX

Companies
Manufacturer
Steris Corporation
2720 Gunter Park East
Montgomery
36109
Alabama
UNITED STATES