Steris 4085 General Surgical Table (2014-07-15)
- Starting date:
- July 15, 2014
- Posting date:
- August 1, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-40817
Recalled Products.
Steris 4085 General Surgical Table
Reason
Steris engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.
Affected products
Steris 4085 General Surgical Table
Lot or serial number
0418009043 to 0417114082
Model or catalog number
ST01XXXX
Companies
- Manufacturer
-
Steris Corporation
2720 Gunter Park East
Montgomery
36109
Alabama
UNITED STATES