Health product recall

Sterile C Section Pack, Bone Marrow Tray, Sterile Labour and Delivery Pack, Sterile Obstetrical Pack, Sterile Delivery Pack, and Bone Marrow Biopsy Pack from Cardinal Health (2018-04-17)

Starting date:
April 17, 2018
Posting date:
May 1, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66642

Affected products

A. STERILE C SECTION PACK

B. BONE MARROW TRAY

C. STERILE LABOUR AND DELIVERY PACK

D. STERILE OBSTETRICAL PACK

E. STERILE DELIVERY PACK

F. BONE MARROW BIOPSY PACK

Reason

Cardinal Health has been advised of an advisory notice on BD Vacutainer Top Tubes contained in these kits.  BD has recently determined that the curing agent thiuram, used in the processing of rubber stoppers since 1996, can lead to release of Carbon Disulfide (CS2) and Carbonyl Sulfide (COS) gases into the headspace of the tubes. These gases can dissolve into the blood sample to chelate lead, and therefore interfere with the accuracy of the Anodic Stripping Voltammetry (ASV) testing methodology. Although the thiuram has been used since at least 1996, the release of CS2 and COS gases into the headspace and their effect on ASV methodology was only recently observed. ASV is the methodology used in magellan diagnostics' leadcare testing systems.

Affected products

A. STERILE C SECTION PACK

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

SMA30CSCLG

SMA30CEMHD

SMA30CSQUL

SMA30CSWEF

SMA30CTCHG

SMA30CSQUJ

SMA30CTCHF

SMA30CSQUK

SOP30BMWPA

Companies
Manufacturer

Cardinal Health

1500 Waukegan Road

Waukegan

60085

Illinois

UNITED STATES



B. BONE MARROW TRAY

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

SOP30BM00C

Companies
Manufacturer

Cardinal Health

1500 Waukegan Road

Waukegan

60085

Illinois

UNITED STATES



C. STERILE LABOUR AND DELIVERY PACK

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

SMA30BILBJ

Companies
Manufacturer

Cardinal Health

1500 Waukegan Road

Waukegan

60085

Illinois

UNITED STATES



D. STERILE OBSTETRICAL PACK

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

SMA30OBPNG

Companies
Manufacturer

Cardinal Health

1500 Waukegan Road

Waukegan

60085

Illinois

UNITED STATES



E. STERILE DELIVERY PACK

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

SMA30DLCLG

SMA30LDRHF

SMA30DECUA

Companies
Manufacturer

Cardinal Health

1500 Waukegan Road

Waukegan

60085

Illinois

UNITED STATES



F. BONE MARROW BIOPSY PACK

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

SBA30BMHMA

Companies
Manufacturer

Cardinal Health

1500 Waukegan Road

Waukegan

60085

Illinois