This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Steri-Vac Gas Sterillizer/Aerator (2014-03-17)

Starting date:
March 17, 2014
Posting date:
April 14, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39019

Recalled Products

Steri-Vac Gas Sterillizer/Aerator

Reason

3M received a single report of a potential exposure OT ethylene oxide (EO) at a hospital installation outside of Canada. The cause was traced to a rare situation involving a failure of the EO cartridge puncture assembly part manufactured between November 10, 2006 and December 14, 2007.

Affected products

Steri-Vac Gas Sterillizer/Aerator

Lot or serial number
  • 350208
  • 350209
  • 350210
  • 350211
Model or catalog number
  • 4XL
  • 5XL 1-DOOR
  • 5XL 2-DOOR
  • 8XL 1-DOOR
  • 8XL 2-DOOR
Companies
Manufacturer

3M Health Care

2510 Conway Ave.

St. Paul

55144

Minnesota

UNITED STATES