StealthStation™ S8 software application
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
StealthStation™ S8 software application |
More than 10 numbers, contact manufacturer. |
9735762 |
Issue
Medtronic has identified the following two software anomalies in StealthStation™ S8 software application versions 1.3.0 and 1.3.2 that can occur within a spine procedure.
1. When lateral orientation radiographic images are taken using a navigation-enabled Ziehm Vision FD Vario 3D C-arm, the navigation orientation information appears flipped when displayed by the StealthStation™ S8 system.
2. When using CT + fluoroscopy images with a slice spacing thickness greater than 2 mm, the system does not display a pop up message indicating the slice spacing is not optimal. Navigation remains accurate in this situation.
Recall start date: August 26, 2024