Stealthstation S8 Application
Brand(s)
Last updated
Summary
Product
Stealthstation S8 Application
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Stealthstation S8 Application | N30873774 N30890766 0011209427 N30873773 | 9735762 |
Issue
Medtronic identified a software anomaly that resulted in the surgical plan data shifting location(s) after the reference exam was changed. For this anomaly to occur, a pre-merge or diffusion-merge must have been used as the initial reference exam.
Recall start date: September 26, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies
Medtronic Navigation, Inc. |
826 Coal Creek Circle, Louisville, Colorado, United States, 80027 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74432