Stealthstation S8 Application
Brand(s)
Last updated
Summary
Product
Stealthstation S8 Application
Issue
Medical devices - Performance issue
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Stealthstation S8 Application |
N30873774 |
9735762 |
Issue
Medtronic identified a software anomaly that resulted in the surgical plan data shifting location(s) after the reference exam was changed. For this anomaly to occur, a pre-merge or diffusion-merge must have been used as the initial reference exam.
Recall start date: September 26, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies
Medtronic Navigation, Inc. |
826 Coal Creek Circle, Louisville, Colorado, United States, 80027 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74432
Get notified