Health product recall

Statclave G4 Chamber Autoclaves

Brand(s)
Last updated

Summary

Product
Statclave G4 Chamber Autoclaves
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Statclave G4 Chamber Autoclaves

710223E00008
710223E00017
710223E00010
710223E00016
710223E00007
710223E00014
710223E00011
710223E00013
710223E00012
710223E00009

G4-622301

Issue

An error in production resulted in missing gaskets in some units on float sensors in the reservoirs. The units were not leaking when they left the facility, however a leak will develop over time with a potential to cause an electrical short.

Recall Start Date: June 16, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Scican Ltd.

1440 Don Mills Road, Toronto, Ontario, Canada, M3B 3P9

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73915

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe