i-STAT Kaolin ACT Cartridge (2017-12-21)
- Starting date:
- December 21, 2017
- Posting date:
- January 12, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65652
Affected products
i-STAT Kaolin ACT Cartridge
Reason
Abbott Point of Care (APOC) has identified that the i-STAT kaolin act test may be prolonged immediately after the administration of an initial full dose aprotinin bolus of 400 kallikrein inhibitor units (kiu)/ml. Once the drug has fully distributed throughout the body there is no impact on the act result reported.
Affected products
i-STAT Kaolin ACT Cartridge
Lot or serial number
More than 1000 numbers, contact manufacturer.
Model or catalog number
03P87-25
Companies
- Manufacturer
-
Abbott Point of Care Inc.
100 and 200 Abbott Park Rd.
Abbott Park
60064
Illinois
UNITED STATES