Health product recall

I-STAT CARTRIDGE (2020-12-22)



Starting date:
December 22, 2020
Posting date:
January 15, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74757


Last updated:

Affected Products 

A. I-STAT CG8+ CARTRIDGE

B. I-STAT EG7+ CARTRIDGE

Reason

Abbott Point of Care has determined through an internal study that the ionized calcium on some i-Stat CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30°C) for periods of time in excess of 3- or 7-days dependent on the cartridge lot number.

Affected products

A. I-STAT CG8+ CARTRIDGE

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

03P88-25

Companies
Manufacturer

Abbott Point of Care Inc.

100 and 200 Abbott Park Rd.

Abbott Park

60064

Illinois

UNITED STATES



B. I-STAT EG7+ CARTRIDGE

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number

03P76-25

Companies
Manufacturer

Abbott Point of Care Inc.

100 and 200 Abbott Park Rd.

Abbott Park

60064

Illinois

UNITED STATES