Stainless Steel Interstitial Needles, 17 Gauge (2019-06-27)
- Starting date:
- June 27, 2019
- Posting date:
- August 2, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70619
Last updated:
Affected Products
Stainless Steel Interstitial Needles, 17 Gauge
Reason
Reports that the trocar tip detached from a 17 gauge stainless steel interstitial needle. The issue has been determined to be caused by a manufacturing process. Lateral force applied to the trocar tip will cause the weld to fragment and the tip to detach.
Affected products
Stainless Steel Interstitial Needles, 17 Gauge
Lot or serial number
R43
S46
T11
T15
T28
T39
T48
U13
Model or catalog number
GM11009500
GM11009510
GM11009520
GM11009530
Companies
- Manufacturer
-
Varian Medical Systems, Inc.
3100 Hansen Way
Palo Alto
94304
California
UNITED STATES