This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

SSP Unitray - High Resolution (2013-11-22)

Starting date:
November 22, 2013
Posting date:
March 25, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38619

Recalled Products

SSP Unitray - High Resolution

Reason

A labeling error was identified where a false negative lead to a mistype. Specifically, a primer mix (R15-05C) indicated a false negative for DRB1*15:34/54/66 alleles. This results in a mistype of DRB1*15:34 as DRB1*16:01:02 and DRB1*15:54/66 as DRB1*15:25.

Affected products

SSP Unitray - High Resolution

Lot or serial number

450102D

450112D

Model or catalog number

More than 10 numbers, contact manufacturer 

Companies
Manufacturer

Life Technologies Corporation (Invitrogen)

9099 North Deerbrook Trail

Brown Deer

53223

Wisconsin

UNITED STATES