SS Retractabed (2014-05-01)
- Starting date:
- May 1, 2014
- Posting date:
- September 26, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-41551
Recalled products
- SS Retractabed
Reason
Insufficient weld was applied to a cross tube that is a support to the intermediate frame of the bed. This cross tube mainly provides strength and stability for the use of a foot side rail if applied. If a foot side rail is applied to bed a weld can fail if not welded correctly. When foot side rail is not applied to bed the chances for failure are reduced.
Affected products
A. SS Retractabed
Lot or serial number
- 20000
- 30000-300191
- 40000-400103
- 50000-50018
- 50000-500342
- 60000-600789
- 70000-701724
Model or catalog number
- SS76
- SS80
Companies
- Manufacturer
-
Med Mizer,
80 Commerce Drive,
Batesville,
99999
Indiana
UNITED STATES