Health product recall

Spot Vision Screener (2018-06-18)

Starting date:
June 6, 2018
Posting date:
October 11, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71223

Last updated:

Affected Products

Spot Vision Screener

Reason

If the device is not used or charged every five months, the coin cell battery that maintains the system clock may become depleted.  This may cause the patient age to be calculated incorrectly. An incorrect patient age could result in an incorrect (either under or over) referral recommendation.

Affected products

Spot Vision Screener

Lot or serial number
  • 732094212228
  • 732094212235
Model or catalog number
  • VS100-B
  • VS100S-B
Companies
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES