This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

SPIRIT ONE BED (2015-07-22)

Starting date:
July 22, 2015
Posting date:
August 10, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54566

Affected Products

A. SPIRIT ONE BED

Reason

Manufacturer received complaints associated with the brakes and scale system. The complaints reported that the brake force may be inadequate or brakes may unexpectedly disengage and that scale readings were inaccurate. Stryker also determined that the design history file of the spirit one bed is incomplete.

Affected products

A. SPIRIT ONE BED

Lot or serial number

All

Model or catalog number
  • B5F00BCFM
  • B5F00C
  • B5F00FM
  • B5F00M
Companies
Manufacturer

Stryker Medical

3800 E Centre Avenue

Portage

49002

Michigan

UNITED STATES