This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Spectrum Infusion Pump with Master Drug Library (2013-09-24)
- Starting date:
- September 24, 2013
- Posting date:
- October 31, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-36485
Recalled Products
Spectrum Infusion Pump with Master Drug Library
Reason
Baxter Corporation is recalling one (1) Sigma Spectrum Infusion Pump with Master Drug Library that was shipped with a component that does not meet standards for withstanding an electrostatic discharge in excess of 15 kilovolts. Such exposure could cause the pump to stop unexpectedly with no audible alarm or visual display of the alarm status.
Affected products
Spectrum Infusion Pump with Master Drug Library
Lot or serial number
- 997333
Model or catalog number
- 35700
Companies
- Manufacturer
-
Baxter Healthcare Corp.
711 Park Ave.
Medina
14103
New York
UNITED STATES