SPECIALTY TRIATHLON TS TRIAL CUTTING GUIDE (2021-03-04)
- Starting date:
- March 4, 2021
- Posting date:
- March 16, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type I
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Healthcare Professionals, General Public, Hospitals
- Identification number:
- RA-75135
Last updated:
Affected Products
SPECIALTY TRIATHLON TS TRIAL CUTTING GUIDE
Reason
Stryker has discovered that the posterior condyle of the Specialty Triathlon TS Trial Cutting Guide(s), sizes 1 through 8, (Left and Right) has the potential to fracture at the augment cut slot. This can occur in the Specialty Triathlon TS Trial Cutting Guide packaging during transport or during impaction of the instrument in surgery. Stryker is aware of complaints associated with the issue. There have been no reports of serious injuries due to the failure mode associated with this recall.
Affected products
SPECIALTY TRIATHLON TS TRIAL CUTTING GUIDE
Lot or serial number
AP15C09
Model or catalog number
I-K3834CG00
Companies
- Manufacturer
-
Howmedica Osteonics Corporation
325 Corporate Drive
Mahwah
07430
New Jersey
UNITED STATES