This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

SPECIALIST 2 INTRAMEDULLARY ROD 400MM (2015-09-09)

Starting date:
September 9, 2015
Posting date:
October 5, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55226

Affected Products

SPECIALIST 2 INTRAMEDULLARY ROD 400MM

Reason

Depuy Orthopaedics is issuing a recall for specific lots of the Specialist 2 Intramedullary (SP2 IM) Rod 400mm Instrument. Further distribution or use of affected lots is to cease immediately. Affected lots manufactured with 455 stainless steel (SS) or 17-4 SS may fracture and leave a portion of the rods in patients.

Affected products

SPECIALIST 2 INTRAMEDULLARY ROD 400MM

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

96-6120

Companies
Manufacturer

Depuy Orthopaedics

Inc.

700 Orthopaedics Drive

Indiana

UNITED STATES