Space Infusion System - Infusomat Space Volumetric Infusion Pump
Brand(s)
Last updated
Summary
Product
Space Infusion System - Infusomat Space Volumetric Infusion Pump
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
---|---|---|
Space Infusion System - Infusomat Space Volumetric Infusion Pump |
More than 10 numbers, contact manufacturer. |
8713051U |
Issue
The pumps have the potential for the upstream occlusion sensor to cease proper function due to the build-up of electrostatic charge during operation triggering false downstream and upstream occlusion alarms.
Recall Start Date: August 28, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
B. Braun Melsungen Ag
Carl-Braun-Str. 1, Melsungen, Germany, 34212
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-74250
Get notified