This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Sonialvision G4

Starting date:
December 30, 2016
Posting date:
January 18, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61838

Reason

The issue is related to the operation for prolonged periods of time at low rate serial exposure, where the heat load capacity of the x-ray anode disc (target) may become damaged and may lead to insulating oil leaking out from the x-ray tube.

Affected products

Sonialvision G4

Lot or serial number

M1C356453010

Model or catalog number

0.7/1.2JG326D-265

Companies

Manufacturer

Shimadzu Corporation

1, Nishinokyo-Kuwabaracho

Nakagyo-Ku

604-8511

JAPAN