Health product recall

Somatom Emotion 16-Slice Configuration and Somatom Perspective - CT Scanner (2018-02-05)

Starting date:
February 5, 2018
Posting date:
March 13, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66124

Affected products

  1. SOMATOM EMOTION 16-SLICE CONFIGURATION
  2. SOMATOM PERSPECTIVE - CT SCANNER

Reason

Siemens healthcare became aware of possible incorrect tube current calculations by the Care Dose4d algorithm for head scans based on p.a. (posterior-anterior) or a.p. topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure.

Affected products

A. SOMATOM EMOTION 16-SLICE CONFIGURATION

Lot or serial number

85828

Model or catalog number

10165877

Companies
Manufacturer

Siemens Shanghai Medical Equipment Ltd.

278 Zhouzhu Road

Shanghai

201318

CHINA



B. SOMATOM PERSPECTIVE - CT SCANNER

Lot or serial number

121001

87596

 

Model or catalog number

10495568

Companies
Manufacturer

Siemens Shanghai Medical Equipment Ltd.

278 Zhouzhu Road

Shanghai

201318

CHINA