Somatom Emotion 16-slice Configuration (2014-06-17)
- Starting date:
- June 17, 2014
- Posting date:
- June 24, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-40191
Recalled Products
- Somatom Emotion 16-slice Configuration
Reason
Siemens has recently detected a rubber component with unfavorable ageing properties on the Somatom Emotion 16-slice Configuration. The mechanical function of this rubber damper, which is situated within the mounting assembly of the rotation motor, may potentially degrade over time and may impact stability.
Affected products
A. Somatom Emotion 16-slice Configuration
Lot or serial number
- S/N 85828
Model or catalog number
- 10165977
Companies
- Manufacturer
-
Siemens Shanghai Medical Equipment Ltd.
278 Zhouzhu Road,
Shanghai
201318
CHINA