SOLTIVE Premium SuperPulsed Laser System
Brand(s)
Last updated
Summary
Product
SOLTIVE Premium SuperPulsed Laser System
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
SOLTIVE Premium SuperPulsed Laser System |
More than 10 numbers, contact manufacturer. |
TFL-PLS |
Issue
Olympus received 98 complaints of sudden power loss or failure to power on in SOLTIVE superpulsed Laser System due to a defect in the 24V power supply. Olympus representatives will begin contacting customers to coordinate the service of affected unit(s) upon establishment of position in the upgrade queue.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Gyrus Acmi, Inc
9600 Louisiana Ave. North, Brooklyn Park, Minnesota, United States, 55445
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81733
Get notified