Solea
Brand(s)
Last updated
Summary
Product
Solea
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Solea | Serial number 0310523 | MODEL 3.0 |
Issue
In normal use, the laser is activated when the user sets the device to the "armed" state and when the foot pedal is depressed in the "armed" state. (the laser is deactivated when it is not in the armed state). However, there were reported events where the laser was activated in the armed state when the user did not actively depress the foot pedal.
Recall start date: August 24, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Dental
Companies
| Convergent Dental, Inc. |
| 140 Kendrick St., Needham, Massachusetts, United States, 02494 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74308
Get notified
Receive emails about new and updated recall and safety alerts.