SoKINOX NO Delivery and Monitoring System (2021-01-25)
- Starting date:
- January 25, 2021
- Posting date:
- February 5, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-74933
Last updated: 2021-02-05
Affected Products
SoKINOX NO Delivery and Monitoring System
Reason
The manufacturer has received a small number of reports regarding leakage due to malfunction of the pilot valve in the backup system. The leakage could in worst case over time drain the connected NO gas cylinders while the system is not in use if the cylinders are left open during that time.
Affected products
SoKINOX NO Delivery and Monitoring System
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
6694550
Companies
- Manufacturer
-
Maquet Critical Care AB
Rontgenvagen 2
Solna
17154
SWEDEN