Sokinox No Delivering And Monitoring System (2018-09-14)
- Starting date:
- September 14, 2018
- Posting date:
- September 28, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67838
Affected products
Sokinox No Delivering And Monitoring System
Reason
Due to a pressure increase after the cylinder switch-over there are 2 possible issues identified:
- The cylinder status icon can change and erroneously indicate remaining gas in an empty NO gas cylinder.
- The alarm "NO cylinder 1 missing (or empty)" or "NO cylinder 2 missing (or empty)" can be deactivated. Since it was found that these pressure creeps were not fully addressed with SW version 1.3, version 1.3.1 was developed which contains additional improvements to further reduce the risk.
Affected products
Sokinox No Delivering And Monitoring System
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
6694550
Companies
- Manufacturer
-
Maquet Critical Care AB
Rontgenvagen 2
Solna
17154
SWEDEN