Health product recall

Softbank II (2013-11-11)

Starting date:
November 11, 2013
Posting date:
January 24, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37653

Recalled products

  1. Softbank II 

Reason

In the Softbank order entry window or the order modify window, keystroke sequence between active and inactive windows while placing a blood product order can cause field validation to fail and the default crossmatch test may be omitted.

Affected products

A. Softbank II

Lot or serial number
  • Used with Softscape GUI versions 1.2.0.X and 1.3.0.X
Model or catalog number

Versions:

  • V.23.1
  • V.23.2
  • V.25.0
  • V.25.1
  • V.25.2
  • V.25.3
Companies
Manufacturer
SCC Soft Computer
5400 Tech Data Drive
Clearwater
33760
Florida
UNITED STATES