This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Softbank II (2013-11-11)

Starting date:
November 11, 2013
Posting date:
January 24, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37653

Recalled products

  1. Softbank II 

Reason

In the Softbank order entry window or the order modify window, keystroke sequence between active and inactive windows while placing a blood product order can cause field validation to fail and the default crossmatch test may be omitted.

Affected products

A. Softbank II

Lot or serial number
  • Used with Softscape GUI versions 1.2.0.X and 1.3.0.X
Model or catalog number

Versions:

  • V.23.1
  • V.23.2
  • V.25.0
  • V.25.1
  • V.25.2
  • V.25.3
Companies
Manufacturer

SCC Soft Computer

5400 Tech Data Drive

Clearwater

33760

Florida

UNITED STATES