Snare, Pediatric, For Slim Gastroscopes
Brand(s)
Last updated
Summary
Product
Snare, Pediatric, For Slim Gastroscopes
Issue
Medical devices - Performance
What to do
| Company has stopped sale, is notifying affected users, and removing the device from sale. |
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Snare, Pediatric, For Slim Gastroscopes |
H04-19-056 |
Not applicable. |
Issue
|
The sheath used for the 7202 device is prone to kinking at the proximal end, against the shrink band, when positioned in a tight loop or tight bend, or when something within the snare loop prevents retraction into the sheath. Recall start date: May 27, 2022 |
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies
Hobbs Medical, Inc.
8 Spring Street, Stafford Springs, Connecticut, United States, 06076
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64245
Get notified
Receive emails about new and updated recall and safety alerts.