Site~Rite 8 Ultrasound System (2019-11-13)
- Starting date:
- November 13, 2019
- Posting date:
- December 6, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71812
Last updated: 2019-12-06
Affected Products
Site~Rite 8 Ultrasound System
Reason
BD has become aware of an issue affecting the Battery Status Indicator on Site~Rite 8 Ultrasound Systems. The indicator may show available charge when the battery is close to depletion. As a result, the Site~Rite 8 has the potential to abruptly shut down, without advance warning to the user, when running on battery power only. This issue will not occur when the unit is plugged into an AC power source, as specified in the Instructions for Use
Affected products
Site~Rite 8 Ultrasound System
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
9770555
Companies
- Manufacturer
-
Bard Access Systems Inc.
605 North 5600 West
Salt Lake City
84116
Utah
UNITED STATES