Reason

BD has become aware of an issue affecting the Battery Status Indicator on Site~Rite 8 Ultrasound Systems. The indicator may show available charge when the battery is close to depletion. As a result, the Site~Rite 8 has the potential to abruptly shut down, without advance warning to the user, when running on battery power only. This issue will not occur when the unit is plugged into an AC power source, as specified in the Instructions for Use