Health product recall

Single Use Oats Set

Brand(s)
Last updated

Summary

Product
Single Use Oats Set
Issue
Medical devices - Device compatibility
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Single Use Oats Set

2003118198
2014118685

AR-8981-08S
AR-8981-06S

Issue

Complaint reports indicate there is a possibility the 8mm reamer in AR-8981-08s may measure 6mm, and the 6mm reamer in AR-8981-06s may measure 8mm. 

Recall start date: November 2, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies

Arthrex Inc

1370 Creekside Boulevard, Naples, Florida, United States, 34108-1945

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72190

Get notified

Receive emails about new and updated recall and safety alerts.