SIGNA™ Explorer
Last updated
Summary
Product
SIGNA™ Explorer
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
SIGNA™ Explorer |
780964EXP |
6433266-1 |
SIGNA Creator |
416431MR |
6433265-1 |
Issue
GE Healthcare became aware that for certain MR systems a scan can be resumed following a power monitor trip when the '6-minute average sar' is above the limit indicated on the 'sar display.'
Recall start date: February 12, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Ge Healthcare (Tianjin) Company Limited
No. 266 Jingsan Road, Tianjin, China, 300308
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76989
Get notified
Receive emails about new and updated recall and safety alerts.