Sigma Spectrum Infusion Pump With Master Drug Library (V8 Platform)
Brand(s)
Last updated
Summary
Product
Sigma Spectrum Infusion Pump With Master Drug Library (V8 Platform)
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Sigma Spectrum Infusion Pump With Master Drug Library (V8 Platform) |
All lots. |
35700BAX2 |
Issue
Baxter is communicating important safety information for Spectrum V8 and Spectrum IQ Infusion Pumps related to potential reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms.
Recall start date: December 29, 2021.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63793
Get notified