Health product recall

Sight Olo Products

Last updated

Summary

Product
Sight Olo Products
Issue
Medical devices - Software defect
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products

Lot or Serial Number

Model or catalog number

Sight Olo-E1

up to and including v2.59.3

OLO-E1

Sight Olo-U1 Hematology Analyzer

up to and including v2.59.3

OLO-U1

Issue

A software defect was discovered, whereby in certain circumstances, a change to a reference range value that was canceled by the operator could nevertheless be saved, which may lead to incorrect reference range being set on the device. The software version of the Olo analyzer will be updated to v2.63.

Recall start date: Jan 18, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Haematology
Companies

S.D. Sight Diagnostics Ltd.

23 Menachem Begin Rd., Tel Aviv

Israel, 6618356

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63876

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe