Shimadzu YSF-300 Fluoroscopy Table (2016-04-15)
- Starting date:
- April 15, 2016
- Posting date:
- May 16, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-58342
Affected products
- Shimadzu YSF-300 Fluoroscopy Table
Reason
Liquid ingress into the control panel could cause a short circuit resulting in the x-ray diagnostic table YSF-300 to move reverse tilting position (trendelenburg position) by itself before the examination was started
Affected products
-
Shimadzu YSF-300 Fluoroscopy Table
Lot or serial number
- 0262A84301
Model or catalog number
- YSF-300
Companies
- Manufacturer
-
Shimadzu Corporation
1, Nishinokyo-Kuwabaracho
Nakagyo-Ku
604-8511
JAPAN