This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Shimadzu YSF-300 Fluoroscopy Table (2016-04-15)

Starting date:
April 15, 2016
Posting date:
May 16, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58342

Affected products

  1. Shimadzu YSF-300 Fluoroscopy Table

Reason

Liquid ingress into the control panel could cause a short circuit resulting in the x-ray diagnostic table YSF-300 to move reverse tilting position (trendelenburg position) by itself before the examination was started

Affected products

  1. Shimadzu YSF-300 Fluoroscopy Table

Lot or serial number
  • 0262A84301
Model or catalog number
  • YSF-300
Companies
Manufacturer
Shimadzu Corporation
1, Nishinokyo-Kuwabaracho
Nakagyo-Ku
604-8511
JAPAN