This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Shandon Rapid Chrome Kwik -Diff Kit

Starting date:
March 22, 2017
Posting date:
April 21, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63062

Reason

The methylene blue, solution #3 in the Kwik-Diff kit was not methylene blue but crystal violet. During investigation, it was verified that the contract manufacturer had labeled the methylene blue solution #3 incorrectly in the kits.

Affected products

Shandon Rapid Chrome Kwik -Diff Kit

Lot or serial number

  • 234
  • 35
  • 36
  • S646
  • S647
  • s645

Model or catalog number

  • 28613
  • 286134

Companies

Manufacturer

RICHARD-ALLAN SCIENTIFIC

SUBSIDIARY OF THERMO FISHER SCIENTIFIC

4481 CAMPUS DRIVE

Kalamazoo

49008

Michigan

UNITED STATES