Sevoflurane, Desflurane, and Isoflurane Electronic Vaporizer Material Package
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Sevoflurane (Quik-Fil) Electronic Vaporizer Material Package |
More than 10 numbers, contact manufacturer. |
115-066755-00 |
Desflurane (Saf-T-Fill) Electronic Vaporizer Material Package |
More than 10 numbers, contact manufacturer. |
115-066759-00 |
Isoflurane (Safety Filling Adapter) Electronic Vaporizer Material Package |
More than 10 numbers, contact manufacturer. |
115-066758-00 |
Issue
The anesthesia system is a device used to administer a general inhalation anesthetic continuously or intermittently to a patient and maintain their ventilation. The A9 anesthesia system utilizes electronic vaporizers (V90) for anesthetic agent delivery.
Mindray has become aware of a potential issue with select batches of V90 electronic vaporizers.
The V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 anesthesia system will generate alarms associated with abnormal vaporizer output. We recommend that the user use the part number reference to identify the license and the device identifier.
Recall start date: July 22, 2024
Additional information
Details
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Mindray Building, Suite 3b, Shenzhen, Guangdong, China, 518057