Health product recall

SERVO-i Ventilator System (2018-03-20)

Starting date:
March 20, 2018
Posting date:
April 13, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66400

Affected products

SERVO-i Ventilator System

Reason

Under certain conditions, an internal communication error might prevent the safety valve to close automatically. As a consequence, ventilation will stop a high priority alarm, as well as clinical alarms, will be triggered. We have no reports of any permanent injuries. This issue will stop ventilation, the safety valve open, thus allowing the patient to breathe freely, but without ventilator support. The high priority technical error alarm TE 11 will be activated simultaneously with other clinical alarms.

Affected products

SERVO-i Ventilator System

Lot or serial number

83775 - 87140

Model or catalog number

64 87 800 E407E

Companies
Manufacturer
Maquet Critical Care AB
Rontgenvagen 2
Solna
17154
SWEDEN