This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

A-Series Ceiling Lift

Starting date:
February 23, 2017
Posting date:
March 16, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62648

Reason

Prism has observed two A-series lift straps break prematurely during internal quality assurance cycle testing at maximum load due to a broken internal magnet that is sewn into the strap.  If the upper magnet breaks it may produce sharp edges that can cut through the strap and reduce its overall life span.

Affected products

A-Series Ceiling Lift

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

324100

324103

324117

324120

324125

Companies

Manufacturer

PRISM MEDICAL LTD

485 MILLWAY AVENUE, UNIT 2,

Concord

L4K 3V4

Ontario

CANADA